Installation
Page 35
USB
Connect a USB cable to computer and run either the Serinus downloader program or an Ecotech
WinAQMS data logger.
TCP/IP (optional)
Plug in a network cable (this cable should be attached to a network) and use the supplied Airodis
software to access the instrument and download data. The Airodis software is supplied on the green
USB resources key provided with this instrument.
Analog/Digital
This port is used to send and receive analog and digital signals. It is normally used to connect with a
gas calibrator or to activate alarm signals.
Each instrument contains 8 digital inputs, 8 digital outputs, 3 analog inputs and 3 analog outputs.
Bluetooth
Connection is enabled using Ecotech’s Serinus remote Android application.
Use the Serinus remote Android application to access instrument and download data. It is available
for download directly from the Google Play Store. Search for “Ecotech Serinus Remote”.
2.3.4 Analyser Set-up
1. Ensure the USB memory key is installed.
2. Check the battery is turned on at the main controller PCB (refer to Figure 5).
3. Turn on the instrument and allow it to warm-up.
4. Check/set time and date (refer to Section 3.5.9).
5. Set the digital filter to the desired monitoring option (refer to Section 3.5.10).
6. Set the internal data logging options (refer to Section 3.5.22).
7. Set the analog/digital inputs and outputs settings (refer to Section 3.5.21).
8. Perform a pressure sensor check (refer to Section 6.4.9).
9. Perform a leak check (refer to Section 6.4.4).
10. Leave the instrument to warm-up and stabilise for 2-3 hours.
11. Perform a multipoint calibration (refer to Section 5.3).
12. The instrument is now ready for operation.