Display Unit iDT-002
MT-91E-A01-B
Page 23
OPERATION
4
4.5 Operation mode
The operation mode provides the operation state, warning, and failure detection.
The operation mode is displayed on the LCD after powering ON the display unit iDT-002.
4.5.1 Operation state
STP pump state is displayed on the upper line of the LCD.
When TMS is enabled, TMS temperature is displayed on the lower line of the LCD.
LCD
Pump operation state
Levitation
Rotor rotation stop.
ACCEL:*****rpm
Acceleration state.
Units of rotational speed are in 500-rpm intervals.
NORM :*****rpm
Normal operation state.
Units of rotational speed are in 500-rpm intervals.
BRAKE:*****rpm
Brake state.
Units of rotational speed are in 500-rpm intervals.
No Levitation
No Levitation state of the rotor.
Autotest
Under initialization at power-up after immediate powering ON
the STP pump.
TMS TEMP:***°C
Current TMS temperature
Table 6 - LCD line