Display Unit iDT-002
MT-91E-A01-B
Page 25
OPERATION
4
4.5.3 Warning contents
When the STP pump detects a warning, a "w" mark appears in the upper left-hand corner of the LCD.
When a "w" is displayed, press the "ENTER" switch to display the warning message. When the LCD
displays the warning message, the display will automatically return to the pump operation state if any
switches are not pressed for two seconds. Or, press the "ENTER" switch to display the pump
operation state immediately.
If more than two warnings are detected, all warnings will be displayed every two seconds. Press the
"ENTER" switch to display the next warning immediately.
Refer to Section 7.2, "WARNING function" in the STP pump Instruction Manual for the details of
warnings.
Levitation
WARNING
1st Damage Limit
B
WARNING
Pump Runtime
C
D
w
A
A.
"w" mark indicating warning detection
B.
Press the "ENTER" switch
C.
No pressing any switches for two seconds or press the "ENTER" switch
D.
No pressing any switches for two seconds or press the "ENTER" switch
Figure 13 - Example of warning indication
Note: The "RESET" switch cannot reset the "WARNING" messages.
Refer to Section 7.2, "WARNING Function," in the STP pump Instruction Manual for the
details of the reset operation.
Note: The STP pump can operate even when a "WARNING" message is displayed.
Overhaul the STP pump when a "WARNING" message is displayed.