Edwards HemoSphere - User Manual

Edwards HemoSphere
238 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
HemoSphere
Advanced Monitor
HemoSphere
Advanced Monitor
Operator’s
manual
Status = Released Printed Date: 2018-01-16 Released Date: 2017-09-14

Table of Contents

Questions and Answers

  • T
    tarachangSep 16, 2025
    What should I do if my Edwards HemoSphere Monitor displays 'Fault: Monitor – Incompatible Software Version'?
    • K
      Kristin FisherSep 16, 2025
      If the Edwards Monitor displays 'Fault: Monitor – Incompatible Software Version', contact Edwards Technical Support, as there was an unsuccessful software upgrade or incompatible software version detected.
  • W
    William MccormickAug 25, 2025
    What to do if the Edwards Monitor displays a 'System Temperature Too High' alert?
    • J
      John CaseyAug 25, 2025
      To address a 'System Temperature Too High' alert on your Edwards Monitor, first, reposition the monitor away from any heat sources. Next, make sure the monitor's ventilation openings are unobstructed and free of dust. If the problem persists, contact Edwards Technical Support.
  • C
    Christopher ClarkAug 25, 2025
    What to do if the Edwards HemoSphere Monitor displays a 'Battery Depleted' fault?
    • J
      Jennifer ArnoldAug 26, 2025
      If your Edwards Monitor displays 'Battery Depleted', connect it to an alternate power source immediately. This will prevent power loss and allow you to continue monitoring.
  • J
    Joy BriggsAug 26, 2025
    What to do if the Edwards Monitor shows an 'Internal System Failure' fault?
    • M
      melissa10Aug 27, 2025
      If the Edwards Monitor displays an 'Internal System Failure' fault, try power cycling the system. If the problem persists, contact Edwards Technical Support.
  • J
    John YoungSep 18, 2025
    What should I do if my Edwards HemoSphere displays 'Fault: Module Slot 1 - Incompatible Software Version'?
    • M
      Melinda CollinsSep 18, 2025
      If the Edwards Monitor displays 'Fault: Module Slot 1 - Incompatible Software Version', contact Edwards Technical Support, as there was an unsuccessful software upgrade or incompatible software version detected.
  • M
    Mark TaylorSep 23, 2025
    What to do if the Edwards Monitor displays 'Alert: System LED Indicators Inoperable'?
    • A
      Ashley HaleySep 23, 2025
      If the Edwards Monitor displays 'Alert: System LED Indicators Inoperable', power cycle the system. If the problem persists, contact Edwards Technical Support.
  • M
    michaeldaviesSep 8, 2025
    How to fix 'Fault: Second Oximetry Cable Detected' on Edwards HemoSphere Monitor?
    • E
      Erica RothSep 8, 2025
      To fix the 'Fault: Second Oximetry Cable Detected' error on your Edwards Monitor, disconnect one of the oximetry cables.
  • A
    Annette SchneiderSep 9, 2025
    What should I do if my Edwards Monitor says 'Fault: Oximetry Cable Disconnected'?
    • H
      holmesrandySep 10, 2025
      If your Edwards Monitor displays 'Fault: Oximetry Cable Disconnected', verify the oximetry cable/catheter connection is secure. Also, check the oximetry cable connector for bent or missing pins.
  • A
    Amy RamirezSep 2, 2025
    What to do if the Edwards HemoSphere Monitor shows 'Fault: Cable Port 2 – Hardware Failure'?
    • C
      Christopher MarshallSep 3, 2025
      If you see a 'Fault: Cable Port 2 – Hardware Failure' message on your Edwards Monitor, try re-inserting the cable. Check for bent or broken pins on the cable or port. As an alternative, try switching to cable port 1. If the problem continues, contact Edwards Technical Support.
  • K
    Kevin PittsSep 1, 2025
    What to do if the Edwards Monitor shows 'Fault: Module Slot 2 – Hardware Failure'?
    • D
      Dustin WoodwardSep 1, 2025
      If you see a 'Fault: Module Slot 2 – Hardware Failure' message on your Edwards Monitor, try reinserting the module. Check for bent or broken pins on the slot or module. As an alternative, try switching to module slot 1. If the problem continues, contact Edwards Technical Support.

Summary

1 Introduction

2 Safety and Symbols

2.2 Warnings

Lists critical warnings for safe operation and potential hazards.

2.3 Cautions

Lists cautions to prevent equipment damage or data invalidation.

3 Installation and Setup

4 HemoSphere Advanced Monitor Quick Start

4.1 HemoSphere Swan-Ganz Module Cardiac Output Monitoring

Brief instructions for setting up and monitoring with the Swan-Ganz module.

4.3 HemoSphere Oximetry Cable Monitoring

Brief instructions for setting up and monitoring with the oximetry cable.

5 Navigating the HemoSphere Advanced Monitor

7 Advanced Settings

7.1 Alarms; Targets

Explains how to configure physiological and technical alarms and target ranges.

9 HemoSphere Swan-Ganz Module Monitoring

9.2 Continuous Cardiac Output

Explains how the monitor measures cardiac output continuously.

10 Monitoring with the HemoSphere Pressure Cable

10.3 FloTrac Sensor Monitoring

Details how to use the FloTrac sensor for monitoring.

11 Oximetry Monitoring

11.2 In Vitro Calibration

Steps for calibrating the oximetry cable before patient use.

13 Troubleshooting

Appendix F: System Care, Service and Support

Edwards HemoSphere Specifications

General IconGeneral
TechnologyAdvanced hemodynamic monitoring
ConnectivityEthernet, USB
Power Supply100–240 VAC, 50/60 Hz
Parameters MeasuredCardiac output, stroke volume, systemic vascular resistance
TrendingYes
AlarmsConfigurable alarms for various parameters
Data StorageInternal memory
ApplicationsCritical care