HemoSphere Advanced Monitor 7 Advanced Settings
NOTE Visual and audible alarm settings are only applicable to parameters being displayed.
To modify Alarms / Targets:
1 Touch the settings icon .
2 Touch Advanced Setup button and enter the required password.
3 Touch Parameter Settings button Alarms / Targets button.
4 Touch anywhere in a parameter box to display the Alarms / Targets popup for the parameter.
NOTE There is a 2 minute inactivity timer associated with this screen.
The red, yellow and green rectangles are fixed shapes, and don’t change size / shape.
7.1.5 Configure All Targets
Targets can easily be configured or changed all at the same time. From the Configure All screen, the user can:
Set Custom Defaults for all parameter alarm and target settings.
Restore all parameter alarm and target settings to Custom Defaults.
Restore all parameter alarm and target settings to Edwards Defaults.
Enable or disable audible alarms for all applicable parameters.
Enable or disable target ranges for all parameters.
1 Touch the settings icon .
2 Touch Advanced Setup button and enter the required password.
3 Touch Parameter Settings button Alarms / Targets button.
Figure 7-1 Alarms / Targets configuration
Status = Released Printed Date: 2018-01-16 Released Date: 2017-09-14