EasyManua.ls Logo

Edwards HemoSphere

Edwards HemoSphere
238 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HemoSphere Advanced Monitor 10 Monitoring with the HemoSphere Pressure Cable
140
3
Use the Select Pressure panel to select the type/location of pressure sensor being used. The
choices for Pressure Transducer are:
ART
CVP
PAP
4 Level the stopcock valve (vent port) just above the TruWave transducer to the patient’s
phlebostatic axis position according to the instructions for use.
5 Open the stopcock valve to measure atmospheric conditions. The pressure should display as a flat
line.
6 Press the physical zero button directly on the pressure cable, or touch the zero button
located on the screen. When zeroing is complete, a tone sounds, and the message “Zero
Complete” appears.
7 Confirm stable zero pressure value and turn stopcocks such that sensors are reading patient
intravascular pressure.
8 If desired, output the pressure signal to a connected patient monitor. See Pressure-Out on page 142
for more information on this option.
9 Touch the home icon to begin monitoring. See table 10-1 for which key parameters are
available based on the type of configuration.
Once pressure cable monitoring is initiated, the blood pressure waveform can also be viewed using the real-
time arterial (ART) waveform display. See Live Arterial Waveform (ART) Display on page 71.
Parameter values monitored using the TruWave DPT are averaged over a 5 second interval, and displayed
every 2 seconds. See table 6-1 on page 95.
10.5 Pressure Cable Monitoring in Swan-Ganz Module Monitoring
Mode
The HemoSphere pressure cable connects to a single Swan-Ganz pulmonary artery pressure port to provide
pulmonary artery pressure (PAP).
While in HemoSphere Swan-Ganz module monitoring mode, the pressure cable can be connected to a
TruWave DPT on a pulmonary artery line.
1 Connect one end of the pressure cable to the HemoSphere advanced monitor.
2 Use a straight in or out motion to connect or disconnect the TruWave DPT. Refer to the TruWave
pressure transducer’s instructions for use and to steps 2-6 in section 10.4.1 above for instructions
on flushing air from the system.
3 Use routine transducer calibration procedures (according to institutional policy) to ensure proper
pressure signals are being transmitted.
4 Touch clinical actions icon More icon Zero & Waveform icon
OR
Press the physical zero button directly on the pressure cable (see figure 10-1).
Status = Released Printed Date: 2018-01-16 Released Date: 2017-09-14

Table of Contents

Related product manuals