190
HemoSphere Advanced Monitor A Specifications
A.2 HemoSphere Advanced Monitor Specifications
HemoSphere
Swan-Ganz
module
Continuous Cardiac Output
(CO), and associated
parameters, both indexed and
non-indexed (SV, SVR, RVEF,
EDV)
Monitors the filament surface temperature and time at temperature. If
a time and temperature threshold is exceeded (above 45 °C),
monitoring halts and alarm triggered.
Measurement of blood temperature within specified accuracy (±0.3
°C). Alarm if blood temperature outside of monitoring range.
Alarm if CO and related parameters outside of alarm ranges. Alarm
delay based on a variable averaging time. Typical averaging time is
57 seconds.
intermittent cardiac output (iCO)
and associated parameters,
both indexed and non-indexed
(SV, SVR)
Measurement of blood temperature within specified accuracy (±0.3
°C).
Alarm if blood temperature outside monitoring range.
HemoSphere
pressure cable
arterial blood pressure (SYS,
DIA, MAP), central venous
blood pressure (CVP),
pulmonary artery blood
pressure (MPAP)
Measurement of blood pressure within specified accuracy (±4% or
±4mmHg, whichever is greater).
Alarm if blood pressure outside of alarm ranges. Alarm delay based
on averaging time is 2 seconds.
The device supports detection of invasive pressure transducer and
transducer cable fault.
The device supports detection of disconnected catheter.
HemoSphere
oximetry cable
oxygen saturation (mixed
venous SvO
2
or central venous
ScvO
2
)
Measurement of oxygen saturation within specified accuracy (±2%
oxygen saturation).
Alarm if oxygen saturation outside of alarm ranges. Alarm delay
based on averaging time is 2 seconds.
Table A-2 HemoSphere advanced monitor physical and mechanical specifications
HemoSphere advanced monitor
Weight 10 ±0.2 lbs (4.5 ±0.1 kg)
Dimensions Height 11.7 in (297 mm)
Width 12.4 in (315 mm)
Depth 5.56 in (141 mm)
Footprint Width 10.6 in (269 mm)
Depth 4.8 in (122 mm)
Display Active Area 12.1 in (307 mm)
Resolution 1024 × 768 LCD
Operating system Windows 7 embedded
Speaker count 1
Table A-1 HemoSphere advanced monitor essential performance – transient and non-transient electromagnetic
phenomena (continued)
Module or
cable
Parameter Essential Performance
Status = Released Printed Date: 2018-01-16 Released Date: 2017-09-14