HemoSphere Advanced Monitor 1 Introduction
23
1.5.2 HemoSphere Pressure Cable
The HemoSphere pressure cable enables vascular pressure
monitoring with a compatible Edwards pressure transducer/
sensor and catheter. A connected FloTrac or FloTrac IQ
sensor provides continuous cardiac output (CO) and
associated hemodynamic parameters. A connected TruWave
transducer provides location based intravascular pressure.
The HemoSphere pressure cable plugs into a monitoring
cable port. For more information, see chapter 10, Monitoring
with the HemoSphere Pressure Cable. Table 1-7 lists the parameters available while using the HemoSphere
pressure cable.
Table 1-6 HemoSphere Swan-Ganz module parameters description
Parameter Description Technology
continuous cardiac output
(CO)
continuous assessment through advanced
thermodilution technology of the volume of
blood pumped by the heart measured in
liters per minute
Swan-Ganz CCO and CCOmbo
catheters
continuous cardiac index
(CI)
continuous cardiac output relative to body
surface area (BSA)
Swan-Ganz CCO and CCOmbo
catheters
intermittent cardiac output
(iCO)
intermittent assessment through the bolus
thermodilution method of the volume of
blood pumped by the heart measured in
liters per minute
Swan-Ganz thermodilution catheters
intermittent cardiac index
(iCI)
intermittent cardiac output relative to body
surface area (BSA)
Swan-Ganz thermodilution catheters
right ventricular ejection
fraction (RVEF)
continuous assessment through advanced
thermodilution technology and algorithm
analysis of the percentage of blood volume
ejected from the right ventricle during
systole
Swan-Ganz CCOmbo V catheters
with ECG signal input
right ventricular end
diastolic volume (EDV)
continuous assessment of the volume of
blood in the right ventricle at the end of
diastole calculated by dividing stroke
volume (mL/beat) by RVEF(%)
Swan-Ganz CCOmbo V catheters
with ECG signal input
stroke volume (SV) amount of blood ejected from the ventricles
with each contraction derived from CO
assessment and heart rate (SV = CO/HR x
1000)
Swan-Ganz CCO,CCOmbo, and
CCOmbo V catheters with ECG
signal input
stroke volume index (SVI) stroke volume relative to body surface area
(BSA)
Swan-Ganz CCO,CCOmbo, and
CCOmbo V catheters with ECG
signal input
systemic vascular
resistance (SVR)
a derived measure of impedance to blood
flow from left ventricle (afterload)
Swan-Ganz CCO and CCOmbo
catheters with MAP and CVP analog
pressure signal input
systemic vascular
resistance index (SVRI)
systemic vascular resistance relative to
body surface area (BSA)
Swan-Ganz CCO and CCOmbo
catheters with MAP and CVP analog
pressure signal input
Status = Released Printed Date: 2018-01-16 Released Date: 2017-09-14