HemoSphere Advanced Monitor 3 Installation and Setup
3.3 HemoSphere Advanced Monitor Installation
3.3.1 Mounting Options and Recommendations
The HemoSphere advanced monitor should be placed on a stable flat surface or securely mounted on a
compatible stand, according to your institution’s practices. The operator should be positioned in front of the
monitor and at close proximity during use. The device is intended to be used by only one user at a time. A
roll stand for the HemoSphere advanced monitor is available as an optional accessory. See Additional
Accessories Description on page 196 more information. Contact your local Edwards representative for
recommendations on additional mounting options.
WARNING Explosion Hazard! Do not use the HemoSphere advanced monitor in the presence
of flammable anesthetic mixture with air or with oxygen or nitrous oxide.
Make sure the HemoSphere advanced monitor is securely positioned or mounted and
that all cords and accessory cables are appropriately arranged to minimize the risk of
injury to patients, users or the equipment.
Do not stack additional equipment or items on top of the HemoSphere advanced
monitor.
The HemoSphere advanced monitor must be positioned in an upright position to
ensure IPX1 ingress protection.
Do not allow any liquids to splash onto the monitoring screen. Liquid buildup may
disable the touchscreen functionality.
Do not position the monitor so that it is difficult to access rear panel ports or power
cord.
Equipment is rated for use with high-frequency surgical equipment. Inaccurate
parameter measurements can be caused by interference from high-frequency surgical
equipment. To reduce hazards that can arise from the use of high-frequency surgical
equipment, only use undamaged patient cables and accessories connected as specified
in this operator’s manual.
This system is rated for use with defibrillators. To ensure proper defibrillator-proof
operation, only use undamaged patient cables and accessories connected as specified in
this operators manual.
All IEC/EN 60950 equipment, including printers, to be positioned no closer than 1.5
meters to the patient’s bed.
CAUTION Do not expose the HemoSphere advanced monitor to extreme temperatures. Refer to
environmental specifications in appendix A.
Do not expose the HemoSphere advanced monitor to dirty or dusty environments.
Do not obstruct the HemoSphere advanced monitor ventilation openings.
Do not use the HemoSphere advanced monitor in environments where strong lighting
makes the LCD screen difficult to view.