184
HemoSphere Advanced Monitor 13 Troubleshooting
13.6.3 SVR Faults/Alert
13.6.4 MPAP Faults/Alert
13.6.5 MAP Faults/Alert
Table 13-15 HemoSphere pressure cable SVR faults/alerts
Message Possible causes Suggested actions
Alert: SVR – Slaved-In CVP
Pressure Signal Loss
HemoSphere advanced monitor analog
input port not configured to accept CVP
Analog input interface cable connection
not detected
Inaccurate input signal
External monitor malfunction
Verify correct voltage range and low/high voltage values
on the HemoSphere advanced monitor for external
monitor
Verify cable connection between the monitoring platform
and bedside monitor is secure
Verify correct height/weight entries and units of measure
for patient’s BSA
Check for signal at external monitor’s analog output
device
Change external device module, if used
Alert: SVR – Configure Analog
Input or Enter CVP for SVR
Monitoring
HemoSphere advanced monitor analog
input port not configured to accept CVP
signal
No CVP value entered
Use the analog input settings screen to configure analog
input ports 1 or 2 for external monitor CVP signal output
Enter CVP value
Table 13-16 HemoSphere pressure cable MPAP faults/alerts
Message Possible causes Suggested actions
Fault: MPAP – Pressure
Disconnected
Pulmonary artery pressure low and non-
pulsatile
Pulmonary artery catheter disconnected
Cable connections not detected
Edwards pressure cable or sensor
malfunction
Internal system malfunction
Verify pulmonary artery catheter connection
Verify connection between pressure cable and sensor and
check for missing pins
Change pressure cable
Change pressure sensor
If problem persists, contact Edwards Technical Support
Table 13-17 HemoSphere pressure cable MAP faults/alerts
Message Possible causes Suggested actions
Fault: MAP – Arterial Pressure
Disconnected
Arterial pressure low and non-pulsatile
Arterial catheter disconnected
Cable connections not detected
Edwards pressure cable or TruWave
sensor malfunction
Internal system malfunction
Verify arterial catheter connection
Verify connection between pressure cable and sensor and
check for missing pins
Change pressure cable
Change pressure sensor
If problem persists, contact Edwards Technical Support