56
HemoSphere Advanced Monitor 4 HemoSphere Advanced Monitor Quick Start
4.2 Monitoring with the HemoSphere Pressure Cable
4.2.1 Pressure Cable Setup
1 Connect the opposite end of the pressure cable to the HemoSphere advanced monitor.
2 Press the power button to turn on the HemoSphere advanced monitor. All functions are accessed
through the touch screen.
3 Select Continue Same Patient button or New Patient button and enter new patient data.
4 Select the Minimally-Invasive monitoring mode button on the Monitoring Mode Selection
window.
5 Touch the home icon .
6 Connect the primed pressure sensor to the pressure cable. The pressure cable LED that surrounds
the zero button at
will flash green indicating that the pressure sensor is detected.
7 Follow all instructions found in pressure monitoring catheter IFU for catheter preparation and
insertion procedures.
The HemoSphere pressure cable must be zeroed before each monitoring session.
4.2.2 Zero Pressure Cable
1 Touch the Zero & Waveform icon located on the navigation bar or through the Clinical
Actions menu.
OR
Press the physical zero button directly on the pressure cable (see figure 4-2).
Figure 4-2 Pressure cable connection overview
pressure sensor/transducer connector
color insert for pressure type
zero button / status LED
HemoSphere pressure cable
HemoSphere advanced monitor
Status = Released Printed Date: 2018-01-16 Released Date: 2017-09-14