HemoSphere Advanced Monitor 9 HemoSphere Swan-Ganz Module Monitoring
4 Connect the compatible Swan-Ganz catheter to the patient CCO cable. See table 9-1 below for
available parameters and required connections.
NOTE Pulmonary artery pressure data is available with a HemoSphere pressure cable
connection. See Pressure Cable Monitoring in Swan-Ganz Module Monitoring Mode on page
140 for more information.
5 Follow the necessary directions for monitoring. See Continuous Cardiac Output on page 120,
Intermittent Cardiac Output on page 123 or EDV/RVEF Monitoring on page 129.
9.1.1 Patient CCO Cable Test
To test the integrity of the Edwards patient CCO cable, perform a cable integrity test. It is recommended to
test the integrity of the cable as part of a troubleshooting process. This does not test the injectate temperature
probe connection of the cable.
Table 9-1 Available HemoSphere Swan-Ganz module parameters and required connections
Parameter Required connection See
CO thermistor and thermal filament connection Continuous Cardiac Output on
page 120
iCO thermistor and injectate (bath or in-line) probe Intermittent Cardiac Output on page 123
EDV/RVEF
(SV)
thermistor and thermal filament connection
*HR slaved-in by HemoSphere advanced monitor
EDV/RVEF Monitoring on page 129
SVR thermistor and thermal filament connection
*MAP and CVP slaved-in by HemoSphere advanced
monitor
SVR on page 132
Status = Released Printed Date: 2018-01-16 Released Date: 2017-09-14