HemoSphere Advanced Monitor 9 HemoSphere Swan-Ganz Module Monitoring
NOTE Appearance of catheters and injectate systems shown in this chapter are for example
only. Actual appearance may vary depending on catheter and injectate system models.
The patient CCO cable and any attached compatible catheter is an APPLIED PART.
1 Ensure that the HemoSphere advanced monitor is off before inserting the HemoSphere Swan-
Ganz module.
2 Insert the HemoSphere Swan-Ganz module into the HemoSphere advanced monitor. The
module will click when properly engaged.
CAUTION Do not force the module into the slot. Apply even pressure to slide and click the module
into place.
3 Press the power button to turn on the HemoSphere advanced monitor and follow steps for
entering patient data. See Patient Data on page 88. Connect the patient CCO cable to the
HemoSphere Swan-Ganz module.
Figure 9-1 HemoSphere Swan-Ganz module connection overview
compatible Swan-Ganz
catheter
thermal filament
connection
thermistor connection
injectate temperature
probe connection
patient CCO cable
HemoSphere Swan-
Ganz module
HemoSphere advanced
monitor
Status = Released Printed Date: 2018-01-16 Released Date: 2017-09-14