HemoSphere Advanced Monitor F System Care, Service and Support
F.2 Cleaning the Monitor and Modules
WARNING Shock or fire hazard! Do not immerse the HemoSphere advanced monitor, modules,
or platform cables in any liquid solution. Do not allow any fluids to enter the
instrument.
The HemoSphere advanced monitor and modules can be cleaned using a lint-free cloth dampened with
cleaning agents that are based on the following chemical content:
• 70% isopropyl alcohol
• 2% glutaraldehyde
• 10% bleach solution (sodium hypochlorite)
• quaternary ammonium solution
Do not use any other cleaning agents. Unless otherwise stated, these cleaning agents are approved for all
HemoSphere advanced monitoring accessories, cables and modules.
CAUTION Do not pour or spray liquid on any portion of the HemoSphere advanced monitor,
accessories, modules, or cables.
Do not use any disinfecting solution other than the types specified.
DO NOT:
Allow any liquid to come in contact with the power connector
Allow any liquid to penetrate connectors or openings in the monitor case or modules
If any liquid does come in contact with any of the above mentioned items, DO NOT
attempt to operate the monitor. Disconnect power immediately and call your
Biomedical Department or local Edwards representative.
F.3 Cleaning the Platform Cables
Platform cables can be cleaned using the cleaning agents listed above in section F.2 and the following
methods.
CAUTION Conduct periodic inspections of all cables for defects. Do not coil cables tightly when
storing.
1 Moisten a lint-free cloth with disinfectant and wipe the surfaces.
2 Follow the disinfectant wipe with rinsing wipes using cotton gauze moistened with sterile water.
Use sufficient rinsing wipes to remove all residual disinfectant.
3 Dry the surface with a clean dry cloth.