HemoSphere Advanced Monitor 6 User Interface Settings
WARNING When switching to a different bedside monitor, always check that the default values
listed are still valid. If necessary, reconfigure the voltage range and corresponding
parameter range or calibrate.
6.2.5.1 Calibration
The calibration option is required when default values are incorrect, or the voltage range is not known. The
calibration process configures the HemoSphere advanced monitor with the analog signal received from the
bedside monitor.
NOTE If the default values are correct, do not calibrate.
CAUTION Only properly trained personnel should calibrate the HemoSphere advanced monitor
analog ports.
1 Touch the settings icon .
2 Touch Monitor Settings button.
3 Touch Analog Input button.
4 Select the desired port number (1 or 2) from the Port list button and corresponding parameter
(MAP or CVP) from the Parameter list button.
5 Select Custom from the voltage value popup screen. The Analog Input Custom Settings screen
will appear.
6 Simulate a full scale signal from the bedside monitor to the selected analog input port on the
HemoSphere advanced monitor.
7 Set the maximum parameter value equal to the full-scale signal value.
8 Touch Calibrate Maximum button. The Maximum A/D value will appear on the Analog
Input Custom Settings screen.
NOTE If an analog connection is not detected, the Calibrate Maximum and Calibrate
Minimum buttons will be disabled and the Maximum A/D value will be displayed as
Not Connected.
9 Repeat the process to calibrate the minimum parameter value.
10 Touch the Accept button to accept the displayed custom settings and return to the Analog Input
Screen.
11 Repeat steps 4-10 to calibrate another port if needed, or touch the home icon to return to
the monitoring screen.