HemoSphere Advanced Monitor 6 User Interface Settings
NOTE While in FloTrac sensor monitoring mode, MAP data through analog input is not
available.
If an analog signal is not detected on the selected port, “Not Connected” will be
displayed below the Port list button.
When an analog input connection or disconnection is first detected, a brief notification
message will be displayed on the status bar.
5 Select CVP in the Parameter list button for the numbered analog port where CVP is connected.
The default setting values for CVP will be displayed.
NOTE The same parameter may not be configured on more than one analog input at the same
time.
While in FloTrac sensor monitoring mode, and a TruWave DPT monitoring CVP is
connected, CVP data through analog input is not available.
6 If the default values are correct for the bedside monitor being used, touch the home icon .
If the default values are not correct for the bedside monitor being used (refer to the bedside
monitor operator’s manual), the user can modify the voltage range, full scale range, or perform the
calibration option described in section 6.2.5.1 of this chapter.
Touch the Full Scale Range value button to change the displayed full scale signal value. Table 6-
2 below shows the allowable input values for full scale range based on the selected parameter.
NOTE A voltage reading of zero is automatically set to a minimum pressure reading of 0
mmHg (0 kPa). The Full Scale Range represents the full scale signal or maximum
pressure reading for the selected Voltage Range.
Touch the Voltage Range list button to change the displayed voltage range. The selectable
voltage ranges available for all of the parameters are:
0 - 1 volts
0 - 5 volts
0 - 10 volts
Custom (see 6.2.5.1: Calibration)
Table 6-2 Analog input parameter ranges
Parameter Full Scale Range
MAP 0 to 510 mmHg (0 kPa to 68 kPa)
CVP 0 to 110 mmHg (0 kPa to 14.6 kPa)