HemoSphere Advanced Monitor 12 Advanced Features
155
parameter color (green/yellow/red). As with other available alarms on the HemoSphere advanced
monitoring platform, the volume of the P(BP) available alarm is adjustable. See Alarms / Targets on
page 100 for information about silencing the alarm and configuring the alarm volume. Occurrence of
P(BP) alarm will be logged in the data download file following an update with P(BP) exceeding the alarm
limit.
12.1.4 P(BP) on Information Bar
When P(BP) is not configured as a key parameter, the parameter value is still computed and displayed in the
information bar as shown in figure 12-3.
12.1.5 Disable P(BP) Information Bar Indicator
To disable the P(BP) information bar indicator:
1 Touch the settings icon .
2 Touch the Parameter Settings button.
3 Touch Advanced Setup button and enter the required passcode.
4 Touch the P(BP) Settings button.
5 Touch the Always alert when P(BP) is high toggle button to switch to Disabled. See
figure 12-4
To re-enable the P(BP) information bar indicator, repeat steps 1-4 and switch the toggle button to Enabled
in step 5.
Figure 12-3 Information bar with P(BP)
Figure 12-4 Parameter settings - P(BP) information bar toggle button
Computed and displayed P(BP) value
Status = Released Printed Date: 2018-01-16 Released Date: 2017-09-14