HemoSphere Advanced Monitor 4 HemoSphere Advanced Monitor Quick Start
3 Touch the clinical actions icon Oximetry Calibration icon .
4 Select Oximetry Type: ScvO
2
or SvO
2
.
5 Touch In vitro Calibration button.
6 Enter either the patient’s hemoglobin (HGB) or hematocrit (Hct) value. A default value may be
used until the patient’s HGB or Hct is available.
7 Touch Calibrate button.
8 When the calibration successfully completes, the following message appears:
In vitro Calibration OK, insert catheter
9 Insert the catheter as described in the catheter directions for use.
10 Touch Start button.
11 If ScvO
2
/SvO
2
are not current key parameters, touch the displayed parameter label located
outside of any parameter globe to select ScvO
2
/SvO
2
as a key parameter from the parameter
popup.
12 Touch within the ScvO
2
/SvO
2
parameter globe to adjust Alarms/Targets.
4.3.2 In vivo Calibration
1 Insert the catheter as described in the catheter directions for use.
2 Insert the optical connector of the catheter “TOP” side up into the oximetry cable and snap the
enclosure shut.
3 Touch clinical actions icon Oximetry Calibration icon .
4 Select Oximetry Type: ScvO
2
or SvO
2
.
5 Touch In vivo Calibration button.
If setup is unsuccessful, one of the following messages will be displayed:
Warning: Wall Artifact or Wedge Detected. Reposition catheter.
OR
Warning: Unstable Signal.
6 If a “Wall Artifact or Wedge Detected,” or “Unstable Signal” message appears, attempt to
troubleshoot the problem as instructed in Chapter 10: Help and Troubleshooting and touch
Recalibrate button to restart the baseline setup.
OR
Touch Continue button to proceed to the Draw operation.
7 When baseline calibration is successful, touch Draw button and then draw the blood sample and
send the blood sample to the lab for measured analysis by co-oximeter.
8 Enter HGB or Hct and ScvO
2
/SvO
2
when lab values are received.
9 Touch Calibrate button.
10 Touch the monitor screen selection icon to select the desired monitoring screen view.
Status = Released Printed Date: 2018-01-16 Released Date: 2017-09-14