HemoSphere Advanced Monitor 8 Data Export and Connectivity Settings
8.4 HIS Connectivity
The HemoSphere advanced monitor has the ability to interface with the Hospital Information
Systems (HIS) to send and receive patient demographics and physiological data. The
HemoSphere advanced monitor supports Health Level 7 (HL7) messaging standard and implements
Integrating Healthcare Enterprise (IHE) profiles. HL7's version 2.6 messaging standard is the most
commonly used means for electronic data exchange in the clinical domain. Use a compatible interface to
access this feature. The HemoSphere advanced monitor HL7 communication protocol, also referred to as
HIS Connectivity, facilitates the following types of data exchanges between the HemoSphere advanced
monitor and external applications and devices:
Sending of physiological data from the HemoSphere advanced monitor to the HIS and/or medical
devices
Sending of physiological alarms and device faults from the HemoSphere advanced monitor to the
HIS
HemoSphere advanced monitor retrieval of patient data from the HIS.
HIS Connection Status should only be queried through the Monitor Settings menu after the HL7
connectivity feature has been configured and tested by the facility network administrator. If HIS Connection
Status is queried while the feature setup is incomplete, the Connection Status Screen will remain open for 2
minutes before timing out.
HIS connectivity status is indicated on the information bar by the symbols shown in Table 8-2.
Figure 8-1 HIS- Patient query screen
Table 8-2 HIS connectivity status
HIS symbol Indication
Connection to all configured HIS actors are good.
Unable to establish communication with configured HIS
actors.
Patient ID is set to “Unknown” in all outbound HIS
messages.
Status = Released Printed Date: 2018-01-16 Released Date: 2017-09-14

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards HemoSphere and is the answer not in the manual?

Edwards HemoSphere Specifications

General IconGeneral
TechnologyAdvanced hemodynamic monitoring
ConnectivityEthernet, USB
Power Supply100–240 VAC, 50/60 Hz
Parameters MeasuredCardiac output, stroke volume, systemic vascular resistance
TrendingYes
AlarmsConfigurable alarms for various parameters
Data StorageInternal memory
ApplicationsCritical care
DisplayTouchscreen

Related product manuals