EasyManua.ls Logo

Edwards HemoSphere

Edwards HemoSphere
238 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
206
HemoSphere Advanced Monitor D Monitor Settings and Defaults
D.4 Alarm and Target Defaults
NOTE Non-indexed ranges are based on indexed ranges and entered BSA values.
Table D-4 Parameter alarm red zone and target defaults
Parameter Units
EW default
lower alarm
(red zone)
setting
EW default
lower target
setting
EW default
upper target
setting
EW default
upper alarm
(red zone)
setting
CI/iCI/sCI
L/min/m
2
1.0 2.0 4.0 6.0
SVI
mL/b/m
2
20 30 50 70
SVRI/iSVRI
dyne-s-m
2
/cm
5
1000 1970 2390 3000
SVV % 0 0 13 20
ScvO
2
/SvO
2
%50657585
EDVI/sEDVI
mL/m
2
40 60 100 200
RVEF/sRVEF %
20 40 60 60
DO
2
ImL
O
2
/min/m
2
300 500 600 800
VO
2
I/VO
2
Ie mL
O
2
/min/m
2
80 120 160 250
CVP mmHg 2 2 8 10
SYS mmHg 90 100 130 150
DIA mmHg 60 70 90 100
MAP mmHg 60 70 100 120
MPAP mmHg 5 9 18 25
HRavg bpm 60 70 100 120
HGB g/dL 7.0 11.0 17.0 19.0
mmol/L 4.3 6.8 10.6 11.8
SpO
2
% 90 94 100 100
PPV % 0 0 13 20
PR bmp 60 70 100 120
P(BP) % 0 0 50 85
Status = Released Printed Date: 2018-01-16 Released Date: 2017-09-14

Table of Contents

Related product manuals