HemoSphere Advanced Monitor 5 Navigating the HemoSphere Advanced Monitor
5.3.1.3 Status Indicators
The lantern at the top of each parameter globe indicates the patient’s current status. The color changes as
the patient’s status changes. The globes may display additional information:
Fault. When a fault condition occurs, the fault message(s) will be displayed on the status bar until the fault
condition is cleared. When there is more than one fault, alert or alarm, the message is cycled every two
seconds.
When a fault condition occurs, parameter calculations are stopped, and each affected parameter globe
displays the last value, time, and date at which the parameter was measured.
Continuous % Change Indicator. This indicator displays the percentage of change, followed by the time
period over which it changed. See Time Intervals / Averaging on page 94 for configuration options.
SVV Filtering Exceeded Indicator. The SVV filtering exceeded indicator symbol appears on the SVV
parameter globe if a high degree of pulse rate variability is detected that could affect the SVV value.
SQI Bar. The SQI bar is a reflection of the signal quality during oximetry monitoring. Signal
quality is based on the catheter condition and positioning within the vessel. For indicator levels, see table 11-
3, “Signal quality indicator levels,” on page 147.
Target Status Indicators. The colored indicator at the top of each monitoring globe indicates the patient’s
clinical status. For indicator colors and their clinical indications, see table 7-2, “Target status indicator
colors,” on page 103.
Figure 5-5 Parameter globe
Audible alarm indicator
– alarms off
Continuous %
Change
Indicator
Target Status
Indicator
Parameter Value
Units
Parameter Name
Additional Symbols
SVV filtering exceeded
indicator (SVV only)
SQI bar
(ScvO
2
/SvO
2
only)
Audible alarm indicator
– alarms paused