194
HemoSphere Advanced Monitor A Specifications
A.5 HemoSphere Pressure Cable Specifications
A.6 HemoSphere Oximetry Cable Specifications
Table A-10 HemoSphere pressure cable specifications
HemoSphere oximetry cable
Weight approximately 0.64 lbs (0.29 kg)
Dimensions Length 15 ft (4.6 m)
Table A-11 HemoSphere pressure cable parameter measurement specifications
Parameter Specification
FloTrac cardiac output (CO) Display range 1.0 to 20 L/min
Reproducibility
1
±6% or 0.1 L/min, whichever is greater
Arterial blood pressure
2
Display range -34 to 312 mmHg
Accuracy ±4% or ±4 mmHg, whichever is greater,
from -30 to 300 mmHg
Bandwidth 1-10Hz
Pulse rate (PR)
Accuracy
3
A
rms
≤3 bpm
1
Coefficient of variation - measured using electronically generated data.
2
Parameter specifications compliant with IEC 60601-2-34 standards. Testing performed under laboratory conditions.
3
Accuracy tested under laboratory conditions.
Table A-12 HemoSphere oximetry cable specifications
HemoSphere oximetry cable
Weight approximately 0.54 lbs (0.24 kg)
Dimensions Length 9.6 ft (2.9 m)
Table A-13 HemoSphere oximetry cable parameter measurement specifications
Parameter Specification
ScvO
2
/SvO
2
Oximetry
(Oxygen Saturation)
Range 0 to 99%
Precision
1
±2% at 30 to 99%
Update rate 2 seconds
1
Precision tested under laboratory conditions.
Status = Released Printed Date: 2018-01-16 Released Date: 2017-09-14