HemoSphere Advanced Monitor 11 Oximetry Monitoring
2 Touch HGB Update button.
3 You can use the displayed HGB and Hct values or touch HGB or Hct buttons to enter a new
value.
4 Touch Calibrate button.
5 To stop the calibration process, touch the cancel icon .
NOTE To achieve optimal accuracy, we recommended you update the HGB and Hct values
when there is a change of 6% or greater in Hct or of 1.8 g/dL (1.1 mmol/L) or greater
in HGB. A change in hemoglobin may also affect SQI. Use HGB Update to resolve
signal quality problems.
11.7 HemoSphere Oximetry Cable Reset
Use HemoSphere oximetry cable reset when the SQI level is continuously high. An oximetry cable reset may
stabilize the signal quality. It should be performed only after attempting other actions to resolve the high SQI
as defined in Troubleshooting.
NOTE The HemoSphere advanced monitor will not permit an oximetry cable reset before
performing a calibration or recalling calibration from the oximetry cable.
1 Touch clinical actions icon Oximetry Calibration icon .
2 Touch Oximetry Cable Reset button.
3 A progress bar will appear. Do not disconnect the oximetry cable.
11.8 New Catheter
Use the New Catheter option any time a new catheter is used for a patient. After New Catheter is
confirmed, oximetry must be re-calibrated Refer to the directions for use provided with each catheter for
specific instructions on catheter placement, calibration type, and use, and for relevant warnings, cautions and
notes.
1 Touch clinical actions icon Oximetry Calibration icon .
2 Touch New Catheter button.
3 Touch Yes button.