HemoSphere Advanced Monitor 6 User Interface Settings
NOTE Connecting to external input devices allows additional information to be displayed. For
example, while monitoring with the HemoSphere Swan-Ganz module and when MAP
and CVP are available continuously from a bedside monitor, SVR is displayed if
configured in a parameter globe. MAP and CVP are displayed on the physio
relationship and physiology monitoring screens.
WARNING The analog communication ports of the HemoSphere advanced monitor share a
common ground that is isolated from the catheter interface electronics. When
connecting multiple devices to the HemoSphere advanced monitor, all devices should
be provided with isolated power to avoid compromising the electrical isolation of any
of the connected devices.
Risk and Leakage current of the final system configuration must comply with IEC
60601-1:2005/A1:2012. It is the responsibility of the user to ensure compliance.
Accessory equipment connected to the monitor must be certified according to IEC/
EN 60950 for data-processing equipment or IEC 60601-1:2005/A1:2012 for electro-
medical equipment. All combinations of equipment must be in compliance with IEC
60601-1:2005/A1:2012 systems requirements.
CAUTION When connecting the HemoSphere advanced monitor to external devices, refer to the
external device’s instruction manual for complete instructions. Verify proper operation
of the system before clinical use.
Once the bedside monitor has been configured for the desired parameter output, connect the monitor via
an interface cable to the selected analog input port on the HemoSphere advanced monitor.
NOTE A compatible bedside monitor must provide an analog output signal.
Please contact your local Edwards representative to obtain the correct HemoSphere
advanced monitor analog input interface cable for your bedside monitor.
The following procedure describes how to configure the analog input ports of the HemoSphere advanced
monitor.
1 Touch the settings icon .
2 Touch Monitor Settings button.
3 Touch Analog Input button.
4 If monitoring with the HemoSphere Swan-Ganz module, select MAP from the Parameter list
button for the numbered analog port where MAP is connected (1 or 2). The default setting values
for MAP will be displayed.
Status = Released Printed Date: 2018-01-16 Released Date: 2017-09-14