HemoSphere Advanced Monitor 6 User Interface Settings
CAUTION The accuracy of continuous SVR while monitoring with the HemoSphere Swan-Ganz
module depends upon the quality and accuracy of the MAP and CVP data transmitted
from the external monitors. Since MAP and CVP analog signal quality from the external
monitor cannot be validated by the HemoSphere advanced monitor, actual values and
the values (including all derived parameters) displayed by the HemoSphere advanced
monitor may not be consistent. The accuracy of continuous SVR measurement,
therefore, cannot be guaranteed. To aid in determining the quality of the analog signals,
regularly compare the MAP and CVP values displayed on the external monitor to the
values displayed on the physio relationship screen of the HemoSphere advanced
monitor. Refer to the external input device operator’s manual for detailed information
regarding accuracy, calibration, and other variables which may impact the analog output
signal from the external monitor.