Hexadecimal Binary Decimal
1 0001 1
2 0010 2
3 0011 3
4 0100 4
5 0101 5
6 0110 6
7 0111 7
8 1000 8
9 1001 9
A 1010 10
B 1011 11
C 1100 12
D 1101 13
E 1110 14
F 1111 15
Each binary digit (bit) represents a ag that is either
acve (state 1) or not acve
(state 0). To help decode the system status word, each bit is numbered (starng with 0
for the least signicant to 31 for the most signicant) as shown below.
Refer to Table: Status ag for the list of the lower 16 status ags that are useful for fault
nding. The upper 16 status ags are reserved by us.
Binary digits
0 0 1 0 0 0 1 0 1 0 0 0 0 0 1 1 0 0 0 0 0 0 0 0 0 0 1 0 0 0 1 0
ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê
3
1
3
0
2
9
2
8
2
7
2
6
2
5
2
4
2
3
2
2
2
1
2
0
1
9
1
8
1
7
1
6
1
5
1
4
1
3
1
2
1
1
1
0
9 8 7 6 5 4 3 2 1 0
Bit numbers
Table 29
Status ag
Bit number
Status ag Acve ags means
0 Fail Fail status condion acve
1 Stopped speed Below stopped speed
2 Normal speed Above normal speed
3 Vent valve closed Vent valve energised
4 Start Start command acve
5 Serial enable Serial enable acve
6 Standby Standby acve
7 Half full speed Above 50% full rotaonal speed
8 Parallel control mode Exclusive control mode selecon
9 Serial control mode Exclusive control mode selecon
04/2022 - ©Edwards Limited
Page 90B80000880_G
B80000880_G - Fault nding