B8J200880_E - Fault finding
Page 113
Fault Yellow service LED is flashing a repeated sequence
Cause A service is required.
Fault The red alarm LED is on
Cause
Normal during the software upgrade process. At other times indicates a problem
has been detected in the FLASH memory.
Fault The red alarm LED is flashing
Cause A fail condition has become active.
Fault Blue bearing LED is flashing a repeated sequence
Cause A bearing service due to inactivity is required.
Fault Blue bearing LED is lit
Cause Bearing run in mode (BRIM) is running.
Fault Pump is not accelerating to full speed
Cause Bearing run in mode (BRIM) is running.
Fault Any other problems
Cause Fault conditions other than the ones previously mentioned.
Remedy Left for RS485, right for RS232.
Right for parallel control and monitoring.
Remedy See Flashing service codes on page 114.
Remedy If during software upgrade then complete the upgrade process. Otherwise cycle
power and if this does not help then perform a software upgrade.
Remedy Note the position of the long flashes within the series of 6 flashes to work out the
error code.
Look up the flashing error code in Flashing error codes on page 114 and follow the
advice given.
Remedy See Flashing inactivity service codes on page 115
Remedy Wait until BRIM is finished. This may take 4.3 hours. We recommend not to cancel
BRIM.
Remedy Wait until BRIM is finished. This may take 4.3 hours. We recommend not to cancel
BRIM, refer to Bearing run in mode on page 121.
Remedy Contact the supplier or us.
11.1 Command set error codes
Error codes that may be returned for serial control and monitoring.
Cause Slide switch is not in the correct position for RS232 or RS485.