B8J200880_E - Configuration
The valve type is now stored in memory within the nEXT pump, but a power cycle is
required before the auxiliary 1 output circuitry adjusts to the stored request.
To check the venting option that is set, send a query as follows:
The reply will be in the following format:
6.1.6 Standby speed setting
The pump can be run at standby speed rather than full rotational speed. The standby
speed is a user-configurable option and can be set to any value between 55% and 100%
full rotational speed.
When the pump is shipped, it is configured with a standby speed of 70% full rotational
speed.
The standby speed setting can be changed using the STDBY+ and STDBY- push buttons
located on the motor controller (shown in Figure Controller view on page 70 ) or by using
a serial command. If using the buttons, each button press will adjust the speed by 15 Hz
(1%).
To change the standby speed setting using a serial command, send the following
command (where the characters 'd' represent the value as a percentage of full rotational
speed):
The reply will be in the following format:
The Standby Speed is now stored in memory within the nEXT pump.
To check what Standby Speed is set, send a query as follows:
The reply will be as follows:
6.1.7 Normal speed setting
The normal speed is a user-configurable setting and can be set to any value between
50% and 100% full rotational speed.
When the pump is shipped, it is configured with a normal speed of 80% full rotational
speed.
To change the normal speed setting, send the following command (where the characters
'd' represent the value as a percentage of full rotational speed):
The reply will be as follows: