EasyManua.ls Logo

Edwards nEXT730 - Earth (Ground) Connections; TIC or TAG Logic Interface Connections; Connect the Logic Interface to Control Equipment; Figure 33 Controller View

Edwards nEXT730
137 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
B8J200880_E - Installation
Page 70
The nEXT pump can be controlled using the Edwards TIC Turbo Instrument Controller,
TIC Turbo Controller or TAG Controller or a customer system. The nEXT pump can be
powered by the Edwards power supply or by a customer power supply via the nEXT
pump power supply cable.
Figure 33 Controller view
1. Bearing LED (indicates when
bearing service is recommended)
2. Standby increase button
3. RS485/CAN/RS232 slide switch 4. USB connector
5. Standby speed decrease button 6. Pump status LEDs
7. Serial receive/transmit LEDs 8. CAN status LED
9. nEXT pump power supply interface 10. Accessory connector
11. nEXT pump logic interface
5.4.1 Earth (ground) connections
Edwards recommend fitting a separate earth (ground) conductor. Refer to Figure:
Connections and controls. The impedance between the pump body and the earth
connection point must be < 0.1 Ω.
5.4.2 TIC or TAG logic interface connections
If an Edwards TIC Turbo Instrument Controller, TIC Turbo Controller or TAG Controller are
used to control the pump, the nEXT pump logic interface cable connects directly into the
back of the TIC or TAG, and the nEXT pump is powered separately using the Edwards
power supply or alternatively the customer power supply connected to the pump via the
dedicated nEXT pump power supply cable.
The RS485/CAN/RS232 slide switch must be in the RS232 position if either the TIC or TAG
is to be used to control the nEXT pump.
5.4.3 Connect the logic interface to control equipment
To operate the pump using your own control system, use suitable connector mating half
(not supplied) to connect the control equipment to the connector on the logic interface
cable.
1
2
3 4
5
9
10 11
8
7
6
Normal
Status
Alarm
Rx
Tx
CAN Status
CS/5036/A

Table of Contents

Related product manuals