B8J200880_E - Fault finding
Page 114
Table 26 Command error codes
Returned error code
Meaning
0
No error
1
Invalid command for object ID
2
Invalid Query/Command
3
Missing parameter
4
Parameter out of range
5
Invalid command in current state - e.g. serial
command to start/stop when in parallel control mode
11.2 Flashing service codes
Whenever a service is required the standard once per revolution flash on the yellow
status LED is replaced with a service flash code.
Table 27 Flashing service codes
Service flash
code
Comments
Actions
LED on 3sec
LED off 1sec
Bearing service recom-
mended
It is recommended that the bearing is
replaced. Refer to Bearing maintenance
on page 103.
LED on 3sec
LED off 3sec
Pump service required
The turbo impeller or controller has
reached its expected life. It is recommen-
ded that the pump is returned to our
service centre for service. Refer to Bearing
maintenance on page 103.
11.3 Flashing error codes
Whenever a fail condition becomes active, the red alarm lights continuously or shows a
flashing sequence.
If the error light is on continuously, this indicates a problem with the embedded
software. In this case, try cycling the power. If cycling the power does not clear the
indiwill be required. Contact the supplier or Edwards. If the alarm LED is flashing,
identify the error flash code and consult the table in Decoding system status words on
page 116.
There is a sufficient off period between each subsequent cycle repetition to clearly mark
the start of a new flash sequence. The duration of a long flash (L) is equal to 3 times the
duration of a short flash (0.5 s).