B8J200880_E - Configuration
Page 82
Abbreviation
Meaning
r
Returned error code - refer to Command set error
codes on page 113
sp
Space character
string
May have several ASCII characters
X
Multi-drop decimal ASCII character
§
*
Fields showing multiple d characters are to indicate typical length. All data fields have
a maximum of 5 decimal characters (prefixed by a minus number for negative numbers).
§
Fields showing multiple X characters are to indicate maximum length and not fixed
length.
Typical setting command:
Command
!
S
8
5
5
sp
d
d
d
cr
Typical setting reply:
Reply

S
8
5
5
sp
r
cr
Typical query command:
Command
?
S
8
5
5
cr
Typical query reply:
Reply
=
S
8
5
5
sp
d
d
d
cr
6.1.3 Power limit setting
The pump is supplied with a default power limit. If this limit is not suitable for the
application, change it to any value between the maximum and minimum setting.
Table 17
Power limit setting
Maximum value
setting
Minimum value
setting
Default power
setting
nEXT730/930
600 W
190 W
500 W
nEXT1230
800 W
300 W
500 W
Send the command as follows (where the 'd' characters represent the value in Watts
that will be set. For example, to set the limit to 500 W, type 500).
Command
!
S
8
5
5
sp
d
d
d
cr
The reply will be in the following format:
Reply

S
8
5
5
sp
r
cr
The Power Limit Setting is now stored in memory within the nEXT pump.
To check what power limit is set, send a query as follows:
Command
?
S
8
5
5
cr
The reply will be in the following format:
Reply
=
S
8
5
5
sp
d
d
d
cr