B8J200880_E - Technical data
Page 50
Object name
Command
Parameter range
Factory
setting
Data type
Units
Comments
6
Hard vent if stop or fail
7
Same as option 6
8
Output is permanently energised (Fan
ON)
9
Do not use
10
Output is permanently de-energised
(Fan OFF)
11
Do not use
12
Do not use
13
Do not use
14
Hard vent when < 50% speed if stop
15
Controlled vent if > 50% speed then
hard vent if < 50% speed if stop.
Bearing Inactivity
(Bearing Condition-
ing)
?V891
0..999999
8760
decimal
hours
Bearing Conditioning period (12
months)
?V891
0..8
0
decimal
-
Bearing Conditioning Status
0 BCON disabled
1 Phase 1 active
2 Phase 2 active
4 Phase 3 active
8 Phase 4 active
!C891
0..1
0
decimal
-
0: BCON Cancel
1: BCON Start
Bearing Inactivity
(Bearing Replace)
?V892
0..999999
26280
decimal
hours
Bearing Replace due to inactivity
threshold period (3 years)