STP-iX455/iXL455 Series Instruction Manual
9-40
Operation
Example:
1.
Remote operation mode : 01
hex
= I/O REMOTE
3.
Vent valve function : FF
hex
= DISABLE
4.
Vent valve starting speed : 32
hex
= 50 %
5.
Vent valve running time : 003C
hex
= 60 sec.
7.
Damage Limit Warning : 00
hex
= ENABLE
8.
Pump Runtime Over Warning : FF
hex
= DISABLE
9.
Pump Runtime Over Warning Hours : 000186A0
hex
=100,000 (x 100 hours)
10.
Imbalance Warning : 00
hex
= ENABLE
12.
Serial communication time out : 003C
hex
= 60 sec.
Parameter
1 2 3 4 5 6 7
ASCII 0
1
*1
*1
*1
*1
F
F
3
2
0
0
3
C
*1
*1
*1
*1
0
0
HEX 30
31
46
46
33
32
30
30
33
43
30
30
Parameter
8 9
10
ASCII F
F
0
0
0
1
8
6
A
0
0
0
HEX 46
46
30
30
30
31
38
36
41
30
30
30
Parameter
11
12
13
ASCII
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
0
0
3
C
*1
*1
*1
*1
*1
*1
HEX
30
30
33
43
Parameter
13
ASCII
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
HEX
*1:
System reservation