STP-iX455/iXL455 Series Instruction Manual
5-2
Installation
Figure 5.1 Attaching Positions of the Baking Heater and Air-cooling Unit
5.2 Cooling the STP Pump
When performing baking or pumping gases, or when operating the STP pump in area of
high humidity or subjected to heat source for unavoidable reason, cool the STP pump.
Use an air-cooling unit (optional accessory), when cooling the STP pump.
The air-cooling unit U0-64 cools only control unit. The air-cooling unit U1-64 cools both
STP pump and control unit. Select the best method which fits your vacuum equipment.
Air-cooling fan
Cooling area
Condition
Baking
Gas pumping
*1
(SCCM)
Environment
U0
-
64
Control unit
N
2
0
to
180
Areas of high
humidity
Areas subjected
to heat source
Ar
0
to
100
U1
-
64
Pump and
control unit
Yes
N
2
0 to 30
0
Ar
0 to 200
*1:
The values shown in the table are typical; they are not guaranteed.
Baking heate
r
Air-cooling unit
(optional accessory)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards STP-iX455 and is the answer not in the manual?

Edwards STP-iX455 Specifications

General IconGeneral
TypeTurbo Molecular Pump
CategoryWater Pump
Outlet PortDN 25 ISO-KF
Frequency50/60 Hz
Pumping Speed455 L/s
Operating Temperature Range5°C to 40°C

Related product manuals