STP-iX455/iXL455 Series Instruction Manual
4-7
Installation
Install the STP pump in a place meeting the following requirements:
Table 4.1 Operating Environment
Ambient Temperature
0 to 40 °C
Ambient Relative Humidity
30 to 95 % (no dew condensing)
Environment
A place free of externally-applied mechanical shock
A place free of a heat source (Keep clear of the heat source or
attach a thermal shield plate)
A place free of a strong magnetic field (Range: up to 15 mT
(150 G) in the axial direction, and up to 3 mT (30 G) in the
radial direction with respect to the rotational axis of the
STP pump)
A place free of a strong electric field
A place free of a source of electric noise
A place free of radiation
No discharge of high voltage (more than 500 V) (If more
than 500 V is discharged, contact Edwards)
A place free of exposure to direct sunlight
A place free of high humidity
A place free of dust
A place free of salty air
A place free of dripping water
A place free of explosive or flammable gas
A place free of corrosive gas
A place free of excessive vibration
STP Pump Installation
Equipment Conditions
Install the STP pump securely so that foreign materials will not
easily fall into the STP pump (Ex.: Si wafers or samples are
positioned above the STP pump) (To prevent foreign materials
from falli
ng into the STP pump, design a shield plate with large
conductance)