Loading...
STP-iX455/iXL455 Series Instruction Manual
9-31
Operation
9.4.9 ReadEvents
Function:
Reads the "Error Record".
It has the recent 10 errors that have been detected.
Transmission Frame:
PCSIM
Stx
0
0
1
? g
Etx
LRC
SIMPC
Ack
PCSIM
Ack
SIMPC
Stx
0
0
1
Sp
g
Parameter 1 to 11
Etx
LRC
Parameter
Item Data format Remarks
1
The number of
"Error Record"
8-bits hexadecimal
coded ASCII
Up to 10 errors
2 to 11
Error Record 1 to
Error Record 10
8-bits hexadecimal
coded ASCII
*1
*1:
Value corresponding to the error is transmitted (see Table 9.9).
The recent error has the smallest parameter number. When the number of errors being
detected is under the maximum number, the value of parameter that is larger than the
number of errors has been detected is set to 0.
Example:
When 3 errors have been detected in the past;
The number of "Error Record" : 03
hex
= 3 errors
Error Record 1 : 0F
hex
= 15 = Disturbance Xb
Error Record 2 : 0D
hex
= 13 = Disturbance Xh
Error Record 3 : 12
hex
= 18 = MOTOR Overheat
Error Record 4 to 10 : 00
hex
= No error recorded
Parameter
1
ASCII 0
3
HEX 30
33
Parameter
2 3 4 5 6 7 8 9 10 11
ASCII 0
F
0
D
1
2
0
0
0
0
0
0
0
0
0
0
0
0
0
0
HEX 30
46
30
44
31
32
30
30
30
30
30
30
30
30
30
30
30
30
30
30

Table of Contents