EasyManua.ls Logo

Edwards STP-iX455

Edwards STP-iX455
147 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
STP-iX455/iXL455 Series Instruction Manual
12-3
Troubleshooting
12.3 WARNING Function Setting
The WARNING functions can be set to Enable or Disable. Set to "ENABLE" when using
the WARNING function. Set to "DISABLE" to release each WARNING function after
the WARNING is detected.
A setting value of the "Pump Run Time Over" is adjustable. The WARNING detection
can be released by setting the value larger than pump running hours after "Pump Run
Time Over" is detected.
See Table 12.2 for the default setting. The setting can be changed via serial
communication, the STP-Link (optional accessory) or the display unit iDT-001 (optional
accessory).
Table 12.2 Default Setting of WARNING Function
WARNING function
Default setting
Remarks
Enable/Disable
Setting
value
Damage Limit ENABLE
Pump Run Time Over DISABLE 0 hour
Variable range: 0 to 30
million hours
(100 hour step).
When the setting value is
0, the WARNING function
is invalid.
Imbalance ENABLE
Refer to Section 9, "Serial Communication Protocol", or the Instruction Manuals
of the STP-Link and display unit for the setting methods.

Table of Contents

Related product manuals