MEASUREMENT PARAMETERS
FH
-
Frequency Limit High
-
Enables entry of frequency limit high (10
MHz
resolution
available). The basic counter and converter will be reset after the data entry.
FL
-
Frequency Limit Low
-.
Enables entry of frequency limit low (10
MHz
resolution
available). The basic counter and conveter will be reset after the data entry.
SELF-TEST FUNCTIONS
TA
-
Test Active
-
Enables the counter to perform the selected test function by entering
the mnemonic TA followed by two digits. When Test 05, 08, 09, or 10 is active and
the counter is being asked to output data, the data that is displayed on the front
panel is the data being output.
The output data format is
as
follows:
XXXXXXXXXXXXCRLF
X
=
alpha-numeric
CR
=
carriage return
LF
=
line feed
For detailed descriptions of tests 01 through 09 and test 11, see the section on Key-
board Controlled Circuit Tests.
Test 10 operates in the following manner:
1. To activate Test 10 input
TA10.
2.
To read the data stored in a specific memory location, input the address of the
memory location in
a
four digit hexadecimal number. Enable the counter to talk
and then read data from the counter.
3.
To alter the data stored in
a
certain memory location:
If
2.
has been-performed
-
input the desired data for that memory location.
If
2.
has not been performed
-
input the memory address, followed by a two
digit hexadecimal number.
TP
-
Test Passive
-
Terminates test function.