Operating faults
BA_Elmasolvex_VA_EN_Vers.05.2018 © Elma Schmidbauer GmbH
• Temperature of the medium in the
media tank, which the indicated partial
step in the error refers to, already too
high? (see temperature table for
Warning 211).
Remedy: use cooled down medium for
the indicated process step.
• Gas content in cleaning medium too
high? Switch off the ultrasound and
basket oscillation temporarily to degas
the cleaning medium more slowly.
• For some programs: Check deflection
and frequency rate oscillation
parameters.
• Switch off cleaning machine and
restart: Contact Service if the fault
115-146
Warning
A fan is not functioning
correctly.
Program is not ended
properly. The cleaning
unit cannot be restarted
while the fault is
present.
A fan is blocked or defective. Check the following possibilities:
• Check fans for foreign bodies and
remove if necessary.
• Switch off cleaning machine and
restart: Contact Service if the fault
persists.
147-178
Warning
Working level filling
level not correct.
Fill level in cleaning chamber
not optimal (or fluctuates
from time to time).
Check the following possibilities:
• Cleaning basket inserted correctly?
• Sieve insert (Fig.9.1.4.1.D) in the
cleaning chamber inserted correctly?
• Fill level in the media tank is too low
(and air is drawn in as a result)?
• Working level fill level sensor and
overfill level fill level sensor
(Fig.9.1.4.1 A/B) soiled?
Clean with suitable cloth if necessary.
• Working level fill level sensor and
overfill level fill level sensor
(Fig.9.1.4.1 A/B) damaged (scratches,
cracks)?
• Quick couplings (Fig. 3.13.1. B/C)
engaged?
• Flat seal in quick coupling (Fig.
4.4.1.4) damaged?
• Gas content in cleaning medium too
high? Switch off the ultrasound and
basket oscillation temporarily to degas
the cleaning medium.
• Switch off cleaning machine and
restart: Contact Service if the fault
persists.