Operating faults
© Elma Schmidbauer GmbH BA_Elmasolvex_VA_EN_Vers.05.2018
• Sieve insert in the cleaning chamber
clogged (Fig.9.1.4.1)?
• Fill level in the media tank is too low
• Can the valves of the quick couplings
(Fig. 4.4.1.2.J) be pressed in / moved
correctly?
• Can the valves of the quick couplings
on the media tanks (Fig. 4.4.1.1.F) be
pressed in / moved correctly?
• Are the quick couplings on the media
tanks (Fig. 3.13.1. B/C) correctly
engaged?
• Suction filter in the media tank (Fig.
3.13.1.G) clogged / blocked?
• Filling level sensor Empty level
(installed under the sieve
insert)(Fig.9.1.4.1.F) soiled? Clean
with suitable cloth if necessary.
• Fill level sensor Empty level
(Fig.9.1.4.1.F) damaged (scratches,
cracks)?
• Switch off cleaning machine and
restart: Contact Service if the fault
215
Abortion
Critical fault!
Venting not possible.
Program has not been
ended.
Vacuum in the cleaning
chamber is not released.
Check the following possibilities:
• Air intake and exhaust openings in the
cleaning chamber (Fig. 3.12.1.A)
clogged?
• Exhaust air tube (chapter 4.3) kinked?
• Switch off cleaning machine and
restart: Contact Service if the fault
persists.
216
Cleaning chamber
drains slowly.
Timeout 1 of the Empty level
fill level sensor during
draining of the cleaning
chamber.
Check the following possibilities:
• Condensate separator, activated
carbon filter (if connected) clogged?
• Exhaust air tube (chapter 4.3) kinked?
• Air intake and exhaust openings in the
cleaning chamber (Fig. 3.12.1.A)
clogged?
• Sieve insert in the cleaning chamber
clogged (Fig.9.1.4.1)?
• Fill level in the media tank is too high
• Can the valves of the quick couplings
(Fig. 4.4.1.2.J) be pressed in / moved
correctly?
• Can the valves of the quick couplings
on the media tanks (Fig. 4.4.1.1.F) be
pressed in / moved correctly?
• Are the quick couplings on the media
tanks (Fig. 3.13.1. B/C) correctly
engaged?