Operating faults
© Elma Schmidbauer GmbH BA_Elmasolvex_VA_EN_Vers.05.2018
91
Sieve insert in the cleaning chamber
clogged (Fig.9.1.4.1)?
Fill level in the media tank is too low
Can the valves of the quick couplings
(Fig. 4.4.1.2.J) be pressed in / moved
correctly?
Can the valves of the quick couplings
on the media tanks (Fig. 4.4.1.1.F) be
pressed in / moved correctly?
Are the quick couplings on the media
tanks (Fig. 3.13.1. B/C) correctly
engaged?
Suction filter in the media tank (Fig.
3.13.1.G) clogged / blocked?
Filling level sensor Empty level
(installed under the sieve
insert)(Fig.9.1.4.1.F) soiled? Clean
with suitable cloth if necessary.
Fill level sensor Empty level
(Fig.9.1.4.1.F) damaged (scratches,
cracks)?
Switch off cleaning machine and
restart: Contact Service if the fault
persists.
215
Abortion
Critical fault!
Venting not possible.
Program has not been
ended.
Vacuum in the cleaning
chamber is not released.
Check the following possibilities:
Air intake and exhaust openings in the
cleaning chamber (Fig. 3.12.1.A)
clogged?
Exhaust air tube (chapter 4.3) kinked?
Switch off cleaning machine and
restart: Contact Service if the fault
persists.
216
Cleaning chamber
drains slowly.
Timeout 1 of the Empty level
fill level sensor during
draining of the cleaning
chamber.
Check the following possibilities:
Condensate separator, activated
carbon filter (if connected) clogged?
Exhaust air tube (chapter 4.3) kinked?
Air intake and exhaust openings in the
cleaning chamber (Fig. 3.12.1.A)
clogged?
Sieve insert in the cleaning chamber
clogged (Fig.9.1.4.1)?
Fill level in the media tank is too high
Can the valves of the quick couplings
(Fig. 4.4.1.2.J) be pressed in / moved
correctly?
Can the valves of the quick couplings
on the media tanks (Fig. 4.4.1.1.F) be
pressed in / moved correctly?
Are the quick couplings on the media
tanks (Fig. 3.13.1. B/C) correctly
engaged?