Fig. 8: air blow through filter housing
Pos: 5.10 /0 020 Über sc hrif ten /ELT RA 1.1 Über sc hri ften /11 Kei ne V erbr en nung EL TRA @ 6\mod_1368524794982_9.docx @ 40132 @ 222 @ 1
3.4 17BNo combustion
Pos: 5.11 /0 025 War nhi nweis e/ G00 02 G EFAH R Le bensg efa hr durc h St rom schl ag - Hochspannung @ 6\mod_1373271950818_9.docx @ 43119 @ 324 @ 1
DANGER
Mortal danger from electric shock
Exposed power contacts - High Voltage
- An electric shock can cause injuries in the form of burns and
cardiac arrhythmia, respiratory arrest or cardiac arrest.
• Disconnect the mains power plug before opening the
analyzer’s cabinet.
Pos: 5.12 /0 010 ELTR A/ 0015 s ervic e_i nstr uc tio ns/C S-800 _S ervic e/ 001 5 F ehler /01 30 M od ul K ei ne V erbren nung F ehler ELT RA @ 6\ mod_1377688599501_9.docx @ 46543 @ 222333333444444 @ 1
After clicking START of analysis or pressing F5 but the sample doesn’t burn, check
the reports in the “status window” on the pc screen.
The following describes how to proceed depending on the displayed message:
3.4.1
84BNo oxygen pressure
If this message permanently remains in the “status window”, then
– Check the oxygen gauge on the analyzer’s front panel.