EasyManuals Logo

ELTRA CS-800 User Manual

ELTRA CS-800
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #93 background imageLoading...
Page #93 background image
93
- Shift the switch with holder downwards until the magnetic switch is off (A
valve clicks and the piston slightly moves downwards).
- Make a second marking on the cylinder (6) at the upper end of the switch
holder (4).
- Adjust the switch holder (4) to have the upper end in the middle of the two
markings.
- Open the furnace and then close it again (F2) to verify the correctness of the
adjustment.
- Adjust the lower cushioning (7), so that the furnace opens without the piston
banging at the bottom.
- Click on “Communication test”-“Digital outputs”, on the 9th button counting
from the left, to energize the coil cooling valve (11/V8), or start analysis
cycle.
- Turn the restrictor (12) counter clockwise until the pressure on the gauge
(13) is reduced by about 0.2 to 0.3 bar (3 to 4 psi) when the coil cooling
valve (11/V8) is open.
- Fix the locknut of the restrictor (12).
Remark:
In case of using valuable gas for operating the pneumatics, the restrictor (12)
should be adjusted for lower flow by turning it clockwise. It can even be completely
closed by turning it clockwise up to the end of its range.
The flow adjusted with this restrictor is for cooling the combustion coil to ensure
very long service life. According to our experience, the combustion coils of our
analyzers don’t need replacement, even after years of operation.
Pos: 7.19 /0 020 Über sc hrif ten /ELT RA 1.1 Über sc hri ften /11 Pres sur e r egul ati on E LTR A @ 6\ mod_1369298860002_9.docx @ 41930 @ 32233 @ 1
4.4 46BPressure regulation
Pos: 7.20 /0 010 ELTR A/ 0015 s ervic e_i nstr uc tio ns/C S-800 _S ervic e/ 002 0 Ei nst ell unge n, T ests u nd F unk tio nsbe schr eib ung /0 255 M odul Pr es sur e Reg ula tio n @ 6\mod_1377688630939_9.docx @ 46975 @ 2424 @ 1
Fig. 48: PR1
PR1 is the inlet oxygen regulator, and is situated inside the analyzer, on the rear
panel.
The outlet pressure needs to be set to 1.5bar (22.5psi).
Pull and rotate the knob of the regulator, until the pressure gauge (6) shows
the correct value.
Ensure that the flow shown on the flow meter (5) shows between 180 and
200l/h; if not, adjust according to the related chapter.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ELTRA CS-800 and is the answer not in the manual?

ELTRA CS-800 Specifications

General IconGeneral
BrandELTRA
ModelCS-800
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals