c) Under Group Registers, select one of the parameters, and then select the right
arrow button (>), or select the double arrow button (>>) to add the whole
group.
8. Enter the update rate. An update rate of for example 10 seconds means that the
plot will be updated every 10 seconds.
9. Select the Start Log button. The logging will proceed until stopped by clicking the
Stop Log button.
Figure 7-23: The Log Registers Function
A. Browse
B. Select register
C. Update rate
D. Start log
7.9.6
Backing up the transmitter configuration
Use this Rosemount Radar Master option to make a backup copy of the configuration
parameters in the transmitter database. The backup file can be used to restore the
transmitter configuration. It can also be used for configuration of a transmitter in a similar
application. Parameters in the saved file can be uploaded directly to the new device.
Procedure
1. In Rosemount Radar Master, select Device Backup Config to File.
2. Browse to the desired directory.
Service and troubleshooting Reference Manual
April 2019 00809-0100-4530
214 Rosemount 5300 Level Transmitter