Procedure
1. In Rosemount Radar Master, enter the service mode.
a) Select Service Enter Service Mode.
b) Type the password and select Submit.
The default password is "admin".
2. Select Service View Input Registers.
3. Select the Dec or Hex checkbox as desired.
Option Description
Dec Display values in decimal format.
Hex Display values in hexadecimal (Hex) format.
4. In the View Input Registers window, do one of the following:
To search for registers by name, select the Names checkbox, and then in the list,
select the desired register group.
To search for registers by entering the register number, select the Numbers
checkbox, and then type the desired register number and the number of
registers to read.
5. Select Read.
7.9.15
View/edit holding registers
Prerequisites
Note
Do not use holding registers to configure the transmitter unless you are qualified. This
dialog is mainly used for service purposes and for advanced configuration.
The holding registers store various transmitter parameters, such as configuration data,
used to control the measurement performance.
Procedure
1. In Rosemount Radar Master, enter the service mode.
a) Select Service Enter Service Mode.
b) Type the password and select Submit.
The default password is "admin".
2. Select Service View Holding Registers.
3. Select the Dec or Hex checkbox as desired.
Option
Description
Dec Display values in decimal format.
Reference Manual Service and troubleshooting
00809-0100-4530 April 2019
Reference Manual 221