FOUNDATION Fieldbus: In the Use `Vapor Compensation´ list, select Yes, and then
select Next.
8. Select the Calibrate Vapor Compensation button and follow the on-screen
instructions.
Note
When the Dynamic Vapor Compensation function has been calibrated, ensure to
keep the Probe Type, Mounting Type, Inner Diameter, Reference Reflector Type,
and remote connection (if used) set as they were during calibration.
If changing the configuration of these parameters, the Dynamic Vapor
Compensation function will be deactivated, and the device error “Vapor
Compensation Not Factory Calibrated” appears. Correct the parameter that has
been changed, and restart the device.
9. When the calibration is finished, restart the device.
a) Select Service Tools → Maintenance.
b) (FOUNDATION Fieldbus only) Select the Reset/Restore.
c) Select Restart Device.
Postrequisites
Verify the echo peak from the reference reflector is marked in the echo curve plot.
Related information
Verify the echo peak from the reference reflector
Verify the echo peak from the reference reflector
Verify that the echo peak from the reference reflector is marked “P2 Reference Reflector”
in the echo curve plot.
Procedure
1. In Rosemount Radar Master, select Setup → Echo Curve.
2. If the echo curve is not read automatically, select Read.
3. Verify that the echo peak from the reference reflector is marked “P2 Reference
Reflector” in the echo curve plot, as illustrated in Figure D-11.
Probes will have the reflector end either at 1.1 ft. (0.35 m) or at 1.6 ft. (0.5 m),
depending on if the probe has a short or a long reflector.