Cerabar S/Deltabar S Description of parameters (ToF Tool)
Endress+Hauser 107
MASS UNIT
Selection
Select mass unit.
Prerequisite:
LIN. MEASURAND = mass
Options:
•g
•kg
•t
•oz
•lb
•ton
•User unit, see also the following parameter description for CUSTOMER UNIT M
and CUST. UNIT FACT. M.
Factory setting:
kg
CUSTOMER UNIT M
Entry
Enter text (unit) for customer-specific mass unit.
You can enter a maximum of eight alphanumeric characters here.
See also CUST. UNIT FACT. M.
Prerequisite:
LIN. MEASURAND = mass, MASS UNIT = user unit
!
Note!
Only the first five characters are shown on the on-site display. For example, if "crates" is
specified as the customer-specific unit, "crate" is displayed.
If the unit contains a slash, up to eight characters can be shown on the on-site display.
The maximum number of characters in the counter is again limited to five. For example,
if "crates/m2" is specified as the customer-specific unit, "crate/m2" is displayed. In the
ToF Tool, all eight characters are displayed.
Factory setting:
_ _ _ _ _ _ _ _
ToF Tool, Table 7: OPERATING MENU SETTINGS BASIC SETUP "Level", LEVEL MODE "Linear"
Parameter name Description