Cerabar S/Deltabar S Description of parameters (ToF Tool)
Endress+Hauser 119
CUSTOMER UNIT V
Entry
Enter text (unit) for customer-specific volume unit.
You can enter a maximum of eight alphanumeric characters here. See also CUST.
UNIT FACT. V
Prerequisite:
COMB. MEASURAND = height and volume, HEIGHT UNIT = user unit
COMB. MEASURAND = %-height and volume,
HEIGHT UNIT = user unit
!
Note!
Only the first five characters are shown on the on-site display. For example, if "crates" is
specified as the customer-specific unit, "crate" is displayed.
If the unit contains a slash, up to eight characters can be shown on the on-site display.
The maximum number of characters in the counter is again limited to five. For example,
if "crates/m2" is specified as the customer-specific unit, "crate/m2" is displayed. In the
ToF Tool, all eight characters are displayed.
Factory setting:
_ _ _ _ _ _ _ _
CUST. UNIT FACT. V
Entry
Enter conversion factor for a customer-specific volume unit.
The conversion factor must be entered in relation to the SI unit "m
3
". See also
CUSTOMER UNIT V.
Prerequisite:
COMB. MEASURAND = height and volume, HEIGHT UNIT = user unit
COMB. MEASURAND = %-height and volume,
HEIGHT UNIT = user unit
Example:
You want the measured value to be displayed in "buckets".
MEASURED VALUE = 0.01 m3 i 1 bucket
Entry CUSTOMER UNIT V: bucket
Entry CUST. UNIT FACT. V: 100
Result: MEASURED VALUE = 1 bucket
Factory setting:
1.0
MASS UNIT
Selection
Select the mass unit for the 2nd measured value.
Prerequisite:
COMB. MEASURAND = height and mass or %-height and mass
Options:
•g
•kg
•t
•oz
•lb
•ton
•User unit, see also the following parameter description for CUSTOMER UNIT M
and CUST. UNIT FACT. M.
Factory setting:
kg