Description of parameters (FF configuration program) Cerabar S/Deltabar S
42 Endress+Hauser
HEIGHT_UNIT_TEXT/
CUSTOMER UNIT H
Selection
Index: 51
Data type: Visible String
Access: Auto, OOS
Enter text (unit) for customer-specific level unit.
You can enter a maximum of eight alphanumeric characters here. → See also
HEIGHT_UNIT_SCALE/CUST. UNIT FACT. H.
Prerequisite:
• MEASURAND_COMBINED = height and volume, HEIGHT_UNIT = user unit
• MEASURAND_COMBINED = height and mass, HEIGHT_UNIT = user unit
• MEASURAND_COMBINED = height and %, HEIGHT_UNIT = user unit
– MEASURAND_COMBINED = COMB. MEASURAND
!
Note!
Only the first five characters are shown on the on-site display. For example, if "crates" is
specified as the customer-specific unit, "crate" is displayed.
If the unit contains a slash, up to eight characters can be shown on the on-site display. The
maximum number of characters in the counter is again limited to five. For example, if
"crates/m2" is specified as the customer-specific unit, "crate/m2" is displayed. In the FF
configuration program, the slash counts as a character, i.e. "crates/m2" would be displayed as
"crates/m".
Prerequisite:
_ _ _ _ _ _ _ _
HEIGHT_UNIT_SCALE/
CUST. UNIT FACT. H.
Entry
Index: 52
Data type: Float
Access: Auto, OOS
Enter conversion factor for a customer-specific level unit.
The conversion factor must be entered in relation to the SI unit "m".
→ See also HEIGHT_UNIT_TEXT/CUSTOMER UNIT H.
Prerequisite:
• MEASURAND_COMBINED = height and volume, HEIGHT_UNIT = user unit
• MEASURAND_COMBINED = height and mass, HEIGHT_UNIT = user unit
• MEASURAND_COMBINED = height and %, HEIGHT_UNIT = user unit
– MEASURAND_COMBINED = COMB. MEASURAND
Example:
– You want the measured value to be displayed in "PU" (PU: packing unit).
– PRIMARY_VALUE = 0.5 m i 1 PU
– Entry HEIGHT_UNIT_TEXT/CUSTOMER UNIT H: PU
– Entry HEIGHT_UNIT_SCALE/CUST. UNIT FACT. H: 2
– Result: PRIMARY_VALUE = 1 PU
Factory setting:
1.0
VOLUME_UNIT/
UNIT VOLUME
Selection
Index: 53
Data type: Unsigned16
Access: Auto, OOS
Select volume unit.
Prerequisite:
• MEASURAND_COMBINED/COMB. MEASURAND = height and volume or
%-height and volume
Options:
•l
•hl
•cm
3
•dm
3
•m
3
•m
3
E
3
•ft
•ft
3
E
3
•gal
•Igal
•bbl
•User unit, → See also VOLUME_UNIT_TEXT/CUSTOMER UNIT V (Page 43) and
VOLUME_UNIT_SCALE/CUST. UNIT FACT. V (Page 43)
Factory setting:
m
3
Pressure Transducer Block
Parameter Description