Cerabar S/Deltabar S Description of parameters (FF configuration program)
Endress+Hauser 41
MEASURAND_
COMBINED/
COMB. MEASURAND
Selection
Index: 48
Data type: Unisgned8
Access: OOS
Select measured variable.
Options:
• Height and volume
• Height and mass
• Height and %
• %-height and volume
• %-height and mass
• %-height and %
!
Note!
This parameter can only be changed if the setting for the
LINEARIZATION parameter (→ Page 36) suits it. For example, this parameter cannot be
changed if the "Flow square root" option was selected for the LINEARIZATION parameter.
Factory setting:
%-height and %
DENSITY_UNIT
Selection
Index: 49
Data type: Unsigned16
Access: Auto, OOS
Select density unit.
Prerequisite:
• MEASURAND_COMBINED = %-height and %, LEVEL_ADJUST_MODE = dry
• MEASURAND_COMBINED = %-height and volume, LEVEL_ADJUST_MODE = dry
• MEASURAND_COMBINED = %-height and mass, LEVEL_ADJUST_MODE = dry
• MEASURAND_COMBINED = height and %, LEVEL_ADJUST_MODE = dry
• MEASURAND_COMBINED = height and volume, LEVEL_ADJUST_MODE = dry
• MEASURAND_COMBINED = height and mass, LEVEL_ADJUST_MODE = dry
– MEASURAND_COMBINED = COMB. MEASURAND
– LEVEL_ADJUST_MODE = CALIBRATION MODE
Options:
• g/cm
3
•kg/dm
3
•kg/m
3
• US lb/in
3
• US lb/ft
3
Factory setting:
kg/dm
3
HEIGHT_UNIT
Selection
Index: 50
Data type: Unsigned16
Access: Auto, OOS
Select level unit.
Prerequisite:
• MEASURAND_COMBINED/COMB. MEASURAND = height and volume, height and
mass or height and %
Options:
•mm
•dm
•cm
•m
•inch
•ft
•User unit, → See also HEIGHT_UNIT_TEXT/CUSTOMER UNIT H (Page 42) and
HEIGHT_UNIT_SCALE/CUST. UNIT FACT. H (Page 42)
Factory setting:
m
Pressure Transducer Block
Parameter Description