Cerabar S/Deltabar S Description of parameters (FF configuration program)
Endress+Hauser 41
MEASURAND_
COMBINED/
COMB. MEASURAND
Selection
Index: 48
Data type: Unisgned8
Access: OOS
Select measured variable.
Options:
Height and volume
Height and mass
Height and %
%-height and volume
%-height and mass
%-height and %
!
Note!
This parameter can only be changed if the setting for the
LINEARIZATION parameter (Page 36) suits it. For example, this parameter cannot be
changed if the "Flow square root" option was selected for the LINEARIZATION parameter.
Factory setting:
%-height and %
DENSITY_UNIT
Selection
Index: 49
Data type: Unsigned16
Access: Auto, OOS
Select density unit.
Prerequisite:
MEASURAND_COMBINED = %-height and %, LEVEL_ADJUST_MODE = dry
MEASURAND_COMBINED = %-height and volume, LEVEL_ADJUST_MODE = dry
MEASURAND_COMBINED = %-height and mass, LEVEL_ADJUST_MODE = dry
MEASURAND_COMBINED = height and %, LEVEL_ADJUST_MODE = dry
MEASURAND_COMBINED = height and volume, LEVEL_ADJUST_MODE = dry
MEASURAND_COMBINED = height and mass, LEVEL_ADJUST_MODE = dry
MEASURAND_COMBINED = COMB. MEASURAND
LEVEL_ADJUST_MODE = CALIBRATION MODE
Options:
g/cm
3
•kg/dm
3
•kg/m
3
US lb/in
3
US lb/ft
3
Factory setting:
kg/dm
3
HEIGHT_UNIT
Selection
Index: 50
Data type: Unsigned16
Access: Auto, OOS
Select level unit.
Prerequisite:
MEASURAND_COMBINED/COMB. MEASURAND = height and volume, height and
mass or height and %
Options:
•mm
•dm
•cm
•m
•inch
•ft
•User unit, See also HEIGHT_UNIT_TEXT/CUSTOMER UNIT H (Page 42) and
HEIGHT_UNIT_SCALE/CUST. UNIT FACT. H (Page 42)
Factory setting:
m
Pressure Transducer Block
Parameter Description