Description of parameters (FF configuration program) Cerabar S/Deltabar S
40 Endress+Hauser
LEVEL_TYPE/
LEVEL MODE
Selection
Index: 45
Data type: Unsigned8
Access: OOS
Select level type.
Options
Linear: the measured variable (level, volume, mass or %) is in direct proportion to the
measured pressure. See also Page 8 ff., Section 4.2.
Pressure linearized: the measured variable (volume, mass or %) is not in direct proportion
to the measured pressure such as in the case of containers with a conical outlet. For the
calibration, enter a linearisation table with at least 2 and not more than 32 points. See
also Page 10 ff., Section 4.3.
Height linearized: select this level type if you require two measured variables or if the
container shape is given with value pairs, e.g. height and volume.
The following combinations are possible:
–Height + volume
–Height + mass
–Height + %
%-height + volume
%-height + mass
%-height + %
Perform two calibrations for this level type. First for the measured variable height or %-
height like for the "Linear" option and then for the measured variable volume, mass or %
like for the "Pressure linearized" option. See also Page 14 ff., Section 4.4.
!
Note!
This parameter can only be changed if the settings for the LINEARIZATION (Page 36) and
PRIMARY_VALUE_TYPE parameters (Page 33) suit it. For example, this parameter cannot
be changed if the "Flow square root" option was selected for the LINEARIZATION
parameter.
Factory setting:
Linear
MEASURAND_LINEAR/
LIN. MEASURAND
Selection
Index: 46
Data type: Unsigned8
Access: OOS
Select measured variable.
Options:
Level
•Volume
•Mass
% (level)
!
Note!
This parameter can only be changed if the setting for the
LINEARIZATION parameter (Page 36) suits it. For example, this parameter cannot be
changed if the "Flow square root" option was selected for the LINEARIZATION parameter.
Factory setting:
% (level)
MEASURAND_
LINEARIZED/
LINd. MEASURAND
Selection
Index: 47
Data type: Unsigned8
Access: OOS
Select measured variable.
Options:
Pressure and volume
Pressure and mass
Pressure and %
!
Note!
This parameter can only be changed if the setting for the
LINEARIZATION parameter (Page 36) suits it. For example, this parameter cannot be
changed if the "Flow square root" option was selected for the LINEARIZATION parameter.
Factory setting:
Pressure and %
Pressure Transducer Block
Parameter Description